|
Next, with SpecView NOT running, find the terminals of the RS485 convertor, they will be labelled in one of several kinds of ways, depending on the manufacturer:
Either the pairs will be labelled A and B, or “+” and “-”
Then the transmit pair will be TD or Tx and the receive pair will be RD or Rx.
So you might find things like: TDA or Tx- or RxB or RD+.
Find the pair that is Transmit (or if there is only one pair, then that’s the one!) and connect the LED/resistor pair onto the terminals. For example, in the above picture, connect the X end to TDA and the Y end to TDB. Please read this note about safety here!
One way around, then LED will be OFF and the other way around the LED will be ON (dimly). If the LED is OFF both ways around, then probably the convertor is not powered, check that with a voltmeter.
Remember which way around the LED needs to be so that it is OFF most of the time.
Now start SpecView and check the Comms Backoff - see here. Go into Runtime mode so that SpecView is trying to communicate to the instruments.
Now when you hold the LED/Resisitor on the terminals, the LED should flash dimly every one or two seconds - this is correct, it shows that SpecView is trying to communicate out of the port. If you get nothing, then the problem is likely to be one of:
- Wrong COM port being used, recheck the settings
- Dead convertor - is there a spare one you can try?
- Dead convertor power supply - check with a voltmeter.
- Convertor does not like a USB-to-Serial convertor
If the LED check shows that SpecView is communicating out of the convertor, you could try to see if the same flashing is present on the terminals of the instrument at the other end of the line. Please read this note about safety here!
|